Online Measurement of Chloride & Sulfate.The 3000CS is a reliable, online analyzer for the direct measurement of chloride and sulfate in pure water and power plant cycle chemistry. This instrument detects these highly corrosive ions to prevent damage to critical power plant components.
Direct Detection of Chloride & Sulfate
The 3000CS provides a method for direct, online detection of corrosive ions without measurement interference from other ions or common additives.
Rapid Monitoring to Prevent Corrosion
Measurement cycles every 45 minutes ensure you can quickly react to chloride and sulfate excursions to prevent corrosive conditions from occurring.
Reduced Monitoring Effort and Costs
Typical chloride and sulfate monitoring requires expensive resources and lab work. The 3000CS eliminates the need for costly offline tests.
Chloride Measurement Range | 0-300 ppb |
Chloride Limit of Detection | 0.5 ppb |
Chloride Measurement Accuracy | ±5% of reading or ± 0.5 ppb whichever is greater |
Sulfate Measurement Range | 0-300 ppb |
Sulfate Limit of Detection | 1 ppb |
Sulfate Measurement Accuracy | |
Measurement Cycle Time | 45 min typical, programmable between 15 minutes and 1 hour |
Sample Flow rate | 25-50 mL/min |
Sample Temperature | 10-45°C (50-113°F) |
Sample Pressure | 0.3-7 bar (5-100 psig) |
Grab Sample Measurement | Included 100 mL capacity |
Power Supply | 100-240 VAC, 50-60 Hz, 100 W typical |
Dimensions (HxWxD) | 927x508x305 mm (36.5"x20.9"x12") |
Weight | 44 kg (97 lbs) |
Ambient Operating Temperature | 10-35°C (50-95°F) |
Humidity | 10-70% non-condensing |
Ratings/Approvals | CE / cULus |
Process Connection | Sample Inlet: 1/4" or 6 mm OD tube SS compressin fitting Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included |
Operator Interface | Color touchscreen Simultaneous display of ion concentrations and analyzer status |
Analog Output | 22mA alarm, 500 ohm max load, not for use with externally powered circuit |
Analog Output Scaling | Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging |
Measurement Parameter | Chloride and sulfate |
Relay Contacts | Mechanical rated at 250 VAC, 3 Amps (Relay 1 NC, Relay 2 to 4 NO), 4-SPDT Type Reed 250 VAC or DC, 0.5 Amps (Relay 5 to 8) |
Analog Output Accuracy | ±0.05 mA |
Model Family | 3000CS |
EnclosureType | Full |
Built-in sequencer | No |
Material Number(s) | 58044001 |