SHIPS AS THE 30139305 as the 30013675 is obsolete and no longer made,
The 6510i ozone sensor with ISM allows users to utilize the full benefits ISM has to offer. These features enable users to maximize the lifetime of the sensors and minimize down time to perform routine maintenance. The Ozone 6510i sensor with ISM operates with the M300 ISM and M800 transmitters.
ISM benefits include:
Full sensor identification
- Sensor Type
- Serial number
Calibration history
- Actual Calibration
- Factory calibration The last 3 calibrations
ACT (Adaptive Calibration Timer)
- A programmable timer to facilitate maintenance planning and increase the operational availability of the measurement loop
- Starting calibration interval: 90 days
TTM (Time To Maintenance for electrolyte & membrane body)
- Ozone concentration
integrated over time to indicate when the membrane and electrolyte need
to be replaced
DLI (Dynamic Lifetime Indicator) for inner body
- Integrates ozone concentration to predict life of inner body and membrane
- The starting values are: Membrane body life time: 180 days; Inner body life time: 1080 days
SAN (Sanitization counter)
- Limit of ozone concentration and duration of sanitization cycle can be defined on the transmitter