Myron L O74FR pH Sensor, 3/4" MNPT, Heavy Duty Flat Tip
All Myron L pH and ORP sensors are combination pH/reference, or ORP/reference. These sensors are designed to operate exclusively with the Myron L 720 Series II pH and ORP monitor/controllers. Each sensor has a uniquely designed, built-in isolated preamplifier that guarantees accurate and reliable measurements - completely eliminating ground-loops and noise issues. The preamp allows for longer distances between the sensor on the monitor/controller without the loss of accuracy or reliability due to cable capacitance, resistance or noise.
The Heavy Duty "F" utilizes a FLAT-TIP self-cleaning sensor (flat glass in place of a round bulb) for use where the most demanding applications are found, such as wastewater. The flat tip will last longer in most abrasive and/or oily solution environments. These sensors utilize a HDPE Double Junction reference with a high temperature - chemical resistant acrylamide gel. Response time, generally 95% in five seconds. This sensor may be installed in ANY direction including Inverted and is simply the best sensor for tough applications.
Specifications:
- Fitting: 3/4" MNPT
- Body Material: CPVC
- Cable Length: 10 foot/3 Meters
- ORP: +/- 2000 mV
- Max Temperature/Max Pressure: 212°F @ 50 psi, 178°F @ 85 psi, 169°F @ 100 psi
- Shelf Life: 6 months