GF Signet 3-2754-1 pH Electrode, Flat, 3K Ohm

Brand: GF Signet

(No reviews yet) Write a Review
MSRP: $658.80
Your Price: $549.00
(You save $109.80 )
Adding to cart… The item has been added
Call For Stock Level
Technical Downloads
GF Signet 3-2754-1 pH Electrode, Flat, 3K Ohm

The Signet 2754-2757 pH and ORP Electrodes feature a unique foul-proof DryLoc® connector with gold-plated contacts designed specifically for use with the Signet 2750 and 2760 Sensor Electronics and preamplifiers. These dependable and highly responsive electrodes feature a large reference chamber volume and intelligent positioning of internal elements that combine to extend the service life. The electrodes are offered with either flat or bulb sensing elements. The flat versions allow sediment and particles to sweep past the measurement surface, minimizing the risks of abrasion, breakage and coating. The bulb versions can be used for general purpose applications and are also offered with a glass or plastic body for use with the 3719 Wet-Tap assembly.

The choice of a PT-1000 or 3KO temperature sensor allows compatibility with all Signet pH/ORP instruments.

The 2754-2757 electrodes are the flagship sensors of the Signet pH/ORP product line. They are all-purpose sensors ideal for a wide range of universal applications.

Features:
- Durable DryLoc® connector with gold plated contacts
- Self-cleaning flat versions reduce risk of glass breakage
- Large reference volume and solid polymer electrolyte
- Temperature sensor included in pH electrodes
- Wet-Tap electrodes available with glass or plastic bodies
- Wet-Tap electrodes can be mounted at any angle

Applications:
- Water & Wastewater Treatment
- Neutralization Systems
- Effluent Monitoring
- Sanitization Systems
- Pool & Spa Control
- Aquatic Animal Life Support Systems
- Process Control
- Cooling Towers
- Boiler Protection

Specifications

2754 thru 2757 pH/ORP Electrodes

General

Compatibility:
     Signet 2750 and/or 2760 Sensor Electronics

Primary Functions:
- 2754 and 2755: Flat surface resists fouling
- 2756 and 2757: Bulb surface for general use
- 2754-HF: Extended use in applications with trace hydrofluoric acid (<2%)
- 2756-DI: Extended use in pure water (< 100 µS)

Operating Range:
- 2754, 2756, 2756-DI : 0 to 14 pH
- 2754-HF: 0 to 12 pH
- 2755, 2757: ± 2,000 mV (ORP)

Temperature Sensor:
- PT-1000 or 3K Balco for pH; none for ORP
- Response time, t = 140 secs. (2754) 196 secs. (2756)

Reference:
- Electrolyte: Solidified Acrylamide Gel
     3.5M KCI
     0.1 M KCl (2756-DI only)
- Element: Ag/AgCl

Efficiency: >97% @ 25°C (77° F)
pH Response Time:
< 5 secs. for 95% of signal change
(for flat and bulb style electrodes)

ORP Response time: Application dependent

Wetted Materials
- Body: CPVC
- Reference Junctions:
     Porous UHMW Polyethylene (standard)
     PTFE (wet-tap)
- Sensing surface: Glass
- O-rings: FPM

Max Temperature/Pressure Rating

Operating Temperature:
- 2756/2756-DI/2757: 0°C to 85°C (32°F to 185°F)
- 2754/2755: 10°C to 85°C (50°F to185°F)
- 2754-HF: 0°C to 50°C (32°F to 122°F)

Maximum Pressure:
- 6.89 bar @ 0°C to 65°C (100 psi @ 32°F to 149°F)
- 4.00 bar @ 65°C to 85°C (58 psi @ 150°F to 185°F)

Storage Temp.: > -10°C (15°F) to 30°C (86°F)
Pipe Size Range:
- 0.5 in. and up: use Signet installation fittings from 0.5 to 4.0 in. (use pipe adapter in pipes over 4 in.)

Mounting:
- Pipe mounting with 2750/2760: =5° to the horizontal plane
- Submersion with 2750/2760: use 3/4 in. NPT or ISO 7-1/R 3/4 in. Male threaded pipe extension

Shipping Weight: 0.25 kg (0.55 lb)
Weight:
2.00 LBS